Rappel de Arrow 4Fr Single Lumen Pressure Injectable Peripherally Inserted Central Catheter (PICC)

Selon New Zealand Medicines and Medical Devices Safety Authority, ce/cet/cette rappel concerne un dispositif en/au/aux/à New Zealand qui a été fabriqué par Arrow International Inc.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    19162
  • Date de mise en oeuvre de l'événement
    2015-10-13
  • Pays de l'événement
  • Source de l'événement
    NZMMDSA
  • URL de la source de l'événement
  • Notes / Alertes
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notes supplémentaires dans les données
    Recalling Organisation: InterMed Medical Ltd, 71 Apollo Drive, Albany, AUCKLAND 1311
  • Cause
    There is a labelling inaccuracy in the instructions for use (ifu) and injection card. the ifu and the injection card contained within the product states that picc testing included 10 pressure injection cycles. this is not correct, it should state that catheter testing included 5 pressure injection cycles., if it were to be used for more than five (5) pressure injections, it cannot be ruled out that there is a potential risk that the catheter could rupture during injections 6-10. catheter rupture can cause catheter or air embolism, bleeding or ineffective treatment and may require the need for radiologic or surgical intervention. no customer complaints have been received for this issue.
  • Action
    Instructions for use to be updated

Device

  • Modèle / numéro de série
    Model: PR-05041-HP, CA-05041-HP, Affected: All lots