Rappel de Carestream DRX Revolution X-Ray System

Selon New Zealand Medicines and Medical Devices Safety Authority, ce/cet/cette rappel concerne un dispositif en/au/aux/à New Zealand qui a été fabriqué par Carestream Health Inc.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    17828
  • Date de mise en oeuvre de l'événement
    2014-12-04
  • Pays de l'événement
  • Source de l'événement
    NZMMDSA
  • URL de la source de l'événement
  • Notes / Alertes
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notes supplémentaires dans les données
    Recalling Organisation: Carestream Health New Zealand Ltd, c/- Quigg Partners, Level 7, 28 Brandon St, Wellington
  • Cause
    The manufacturer has received a report that an overseas service engineer observed a loose boom arm during routine maintenance on a drx revolution mobile x-ray system as one of three bolts that secure the boom to the carriage had fractured and the other two bolts were loose causing the boom arm to become loose. no other reports of this type have been received since the device was introduced into commercial distribution in 2009.
  • Action
    Product to be modified

Device

  • Modèle / numéro de série
    Model: , Affected: Manufactured prior to 15-Oct-2013