Rappel de Draeger Monitor AC3000 Spring Arm

Selon New Zealand Medicines and Medical Devices Safety Authority, ce/cet/cette rappel concerne un dispositif en/au/aux/à New Zealand qui a été fabriqué par Draegerwerk AG & Co KGaA.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    15702
  • Date de mise en oeuvre de l'événement
    2013-11-21
  • Pays de l'événement
  • Source de l'événement
    NZMMDSA
  • URL de la source de l'événement
  • Notes / Alertes
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notes supplémentaires dans les données
    Recalling Organisation: Draeger New Zealand Ltd, Unit 4, 24 Bishop Dunn Place, East Tamaki, Auckland 2013
  • Cause
    Manufacturer has become aware of a case in which the monitor or support bracket has detached from the joint pin at the bottom of the height adjustable ac3000 spring arm and fallen down together with the monitor until it was only held by the electrical cable connections. no one was injured but a falling monitor support has the potential to cause personal injury if the event was repeated. an inspection of a further seven monitor brackets in the same hospital identified severe wear on the ring groove responsible for the joint pin.
  • Action
    Product to be modified