Avis De Sécurité sur Artis zee systems

Selon Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), ce/cet/cette avis de sécurité concerne un dispositif en/au/aux/à Slovenia qui a été fabriqué par Siemens Healthcare GmbH.

Qu'est-ce que c'est?

Les avis de sécurité sont des communications que les fabricants de dispositifs médicaux ou leurs représentants font parvenir en particulier aux travailleurs du secteur médical, mais aussi aux utilisateurs de ces dispositifs. Les avis de sécurité concernent les actions entreprises quant aux dispositifs présents sur le marché et peuvent inclure des rappels et des alertes.

En savoir plus sur les données ici
  • Type d'événement
    Field Safety Notice
  • Date
    2016-02-15
  • Pays de l'événement
  • Source de l'événement
    AMPMDRS
  • Notes / Alertes
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notes supplémentaires dans les données
  • Cause
    This corrective action addresses two possible, mutually independent causes of a system defect. o in artis zeego systems, angulations in the vicinity of the c-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. 0 in artis systems with a100plus generators, an attempt to resume operation following detection of a fault (such as a short circuit in the x-ray tube) can result in the failure of a module in the high-voltage generator. what is the imgact on system operation and what is the potential risk? a fault can result in a spontaneous system malfunction that can only by rectified by our service organization.

Device

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    AMPMDRS