Avis De Sécurité sur CARDIOVIT AT-102 Plus

Selon Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), ce/cet/cette avis de sécurité concerne un dispositif en/au/aux/à Slovenia qui a été fabriqué par SCHILLER.

Qu'est-ce que c'est?

Les avis de sécurité sont des communications que les fabricants de dispositifs médicaux ou leurs représentants font parvenir en particulier aux travailleurs du secteur médical, mais aussi aux utilisateurs de ces dispositifs. Les avis de sécurité concernent les actions entreprises quant aux dispositifs présents sur le marché et peuvent inclure des rappels et des alertes.

En savoir plus sur les données ici
  • Type d'événement
    Field Safety Notice
  • Date
    2017-01-09
  • Pays de l'événement
  • Source de l'événement
    AMPMDRS
  • Notes / Alertes
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notes supplémentaires dans les données
  • Cause
    Patient data mix-up may happen when the setting of “manual send" and “automatic erase after send” is chosen for the data transmission. ecg data interchange only happens with this particular setting. all other settings are not affected. the occurrence is evaluated as rare, considering the limited number of incident reports in relation to the number of ecgs made per year and due to the fact that most customers use auto transmission, and additionally to the fact that data interchange is not likely to happen if the patient is released from the cables before the manual ecg transmission is started.