Avis De Sécurité sur Cios Alpha mobile C-Arm system

Selon Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), ce/cet/cette avis de sécurité concerne un dispositif en/au/aux/à Slovenia qui a été fabriqué par Siemens Healthcare GmbH.

Qu'est-ce que c'est?

Les avis de sécurité sont des communications que les fabricants de dispositifs médicaux ou leurs représentants font parvenir en particulier aux travailleurs du secteur médical, mais aussi aux utilisateurs de ces dispositifs. Les avis de sécurité concernent les actions entreprises quant aux dispositifs présents sur le marché et peuvent inclure des rappels et des alertes.

En savoir plus sur les données ici
  • Type d'événement
    Field Safety Notice
  • Pays de l'événement
  • Source de l'événement
    AMPMDRS
  • Notes / Alertes
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notes supplémentaires dans les données
  • Cause
    1) if the system language is set to english and the yyyymmdd date format 5 used, the patient‘s date of birth received through the dicom worklist and query/retrieve will be displayed with month and day reversed for patients who have birthdays on days 1 through 12. birth dates for days 13 through 31 are displayed correctly. the issue only appears to happen for days 1 through 12. 2) system crash may occur during dvd live recording if radiation is released when the dvd icon on the flc status bar is blinking. in this case error messages wili be displayed and a system reset is necessary. 3) sporadically the system will not be ready for radiation during reconnection of the trolley to the main unit. a system reboot is necessary.

Device

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    AMPMDRS