Avis De Sécurité sur Foot Switch - AngioJet™ Ultra 5000A Console

Selon Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), ce/cet/cette avis de sécurité concerne un dispositif en/au/aux/à Slovenia qui a été fabriqué par Boston Scientific International S.A..

Qu'est-ce que c'est?

Les avis de sécurité sont des communications que les fabricants de dispositifs médicaux ou leurs représentants font parvenir en particulier aux travailleurs du secteur médical, mais aussi aux utilisateurs de ces dispositifs. Les avis de sécurité concernent les actions entreprises quant aux dispositifs présents sur le marché et peuvent inclure des rappels et des alertes.

En savoir plus sur les données ici
  • Type d'événement
    Field Safety Notice
  • Date
    2017-02-07
  • Pays de l'événement
  • Source de l'événement
    AMPMDRS
  • Notes / Alertes
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notes supplémentaires dans les données
  • Cause
    Boston scientific has received field reports regarding the angiojet ultra 5000a console foot switch sticking during use resulting in an undesired continued operation of the device. if this situation occurs, perform the troubleshooting steps per the operator’s manual. an attempt can be made to manually release the footswitch, or turn the console off to deactivate the catheter. the complaint review shows that this issue results in a prolongation of the procedure in order to perform troubleshooting steps and remove the catheter. the most serious consequence from this issue could be the need to abort a procedure if the situation has not been resolved and alternative options are not available. to date, there have been no reported patient adverse events as a result of these complaints. .

Manufacturer

  • Société-mère du fabricant (2017)
  • Commentaire du fabricant
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.”
  • Source
    AMPMDRS