Avis De Sécurité sur neoBLUE® blanket system fiber optic pads

Selon Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), ce/cet/cette avis de sécurité concerne un dispositif en/au/aux/à Slovenia qui a été fabriqué par Natus Medical.

Qu'est-ce que c'est?

Les avis de sécurité sont des communications que les fabricants de dispositifs médicaux ou leurs représentants font parvenir en particulier aux travailleurs du secteur médical, mais aussi aux utilisateurs de ces dispositifs. Les avis de sécurité concernent les actions entreprises quant aux dispositifs présents sur le marché et peuvent inclure des rappels et des alertes.

En savoir plus sur les données ici
  • Type d'événement
    Field Safety Notice
  • Pays de l'événement
  • Source de l'événement
    AMPMDRS
  • Notes / Alertes
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notes supplémentaires dans les données
  • Cause
    Natus medical has received feedback from some customers about early failure of the neoblue® blanket system fiber optic pads. these failures involve discoloration/degradation followed by eventual melting of the fiber optic bundle at the connector that is inserted in the neoblue blanket light box. natus’ investigations have shown that this failure occurs after extended exposure to the intense light source within the box, at which time the pad no longer provides the therapeutic treatment for which it is intended. following the guidelines of this technical bulletin will help to prevent failures from occurring on customer systems. .

Device

Manufacturer