Avis De Sécurité sur Various Sterile-Packaged Knee and Hip Implants

Selon Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), ce/cet/cette avis de sécurité concerne un dispositif en/au/aux/à Slovenia qui a été fabriqué par Zimmer Biomet.

Qu'est-ce que c'est?

Les avis de sécurité sont des communications que les fabricants de dispositifs médicaux ou leurs représentants font parvenir en particulier aux travailleurs du secteur médical, mais aussi aux utilisateurs de ces dispositifs. Les avis de sécurité concernent les actions entreprises quant aux dispositifs présents sur le marché et peuvent inclure des rappels et des alertes.

En savoir plus sur les données ici
  • Type d'événement
    Field Safety Notice
  • Date
    2016-08-08
  • Pays de l'événement
  • Source de l'événement
    AMPMDRS
  • Notes / Alertes
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notes supplémentaires dans les données
  • Cause
    Zimmer biomet is initiating a recall of specific lots of foil pouched, sterile trabecular metal implants that were manufactured between april 2011 and october 2012. the foil pouches are part of a dual barrier system to provide sterile integrity. retrospective testing performed as a result of an internal review of the packaging system revealed that in some instances, tears/holes in some of the foil pouches were present in either the inner foil pouch or the outer foil pouch, but never both in the same sample. no product complaints have been reported for this issue. zimmer biomet tmt implemented a packaging change from the foil pouch sterile barrier system to a double sterile barrier thermoform cavity configuration by november 2012. the affected product was distributed between april 2011 and march 2016. below are examples of failures seen during this testing. .

Device

Manufacturer

  • Société-mère du fabricant (2017)
  • Commentaire du fabricant
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.