Rappel de Sterilized gowns and towels. || C-Section Pack - contains 8 towels, 1 each large single-ply scrub nurse gown/towel, 2 each large gortex surgeons gown/towel, 2 each 36'' re-usable wrappers, 1 24'' re-usable wrapper. || Surgical gown packs, inside double wrappers, for use in surgery and operating rooms. Packs include gowns and towels. || Packages are labeled ''STERILE SURGICAL SYSTEMS, LLC ** Tumwater, Washington ** Catalog # ** Quantity ** MFG Date ** Order No. ** Contents ** Contents of the unopened, undamaged package are sterile. Do not resterilize***.

Selon U.S. Food and Drug Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à United States qui a été fabriqué par Sterile Surgical Systems Llc.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    25118
  • Classe de risque de l'événement
    Class 2
  • Numéro de l'événement
    Z-0441-03
  • Date de mise en oeuvre de l'événement
    2002-06-04
  • Date de publication de l'événement
    2003-01-14
  • Statut de l'événement
    Terminated
  • Pays de l'événement
  • Date de fin de l'événement
    2003-01-30
  • Source de l'événement
    USFDA
  • URL de la source de l'événement
  • Notes / Alertes
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notes supplémentaires dans les données
    Gown, Surgical - Product Code FYA
  • Cause
    An indicator inside the package did not change to indicate gowns were sterilized.
  • Action
    The firm telephoned it's two customers on 6/4/02 and offered to have product returned or provide documentation that the sterlizing indicators were defective, and theoretically the gown packs were sterilized.

Device

  • Modèle / numéro de série
    MFG Dates: 5/24/02, 5/28/02, 5/29/02, 5/30/02, 5/31/02, 6/3/02
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    The firm distributed to two hospitals located in the greater Seattle, WA area.
  • Description du dispositif
    Sterilized gowns and towels. || C-Section Pack - contains 8 towels, 1 each large single-ply scrub nurse gown/towel, 2 each large gortex surgeons gown/towel, 2 each 36'' re-usable wrappers, 1 24'' re-usable wrapper. || Surgical gown packs, inside double wrappers, for use in surgery and operating rooms. Packs include gowns and towels. || Packages are labeled ''STERILE SURGICAL SYSTEMS, LLC ** Tumwater, Washington ** Catalog # ** Quantity ** MFG Date ** Order No. ** Contents ** Contents of the unopened, undamaged package are sterile. Do not resterilize***.
  • Manufacturer

Manufacturer

  • Adresse du fabricant
    Sterile Surgical Systems Llc, 2757 29TH AVENUE S.W., Tumwater WA 98512
  • Source
    USFDA