Philips Medical Systems North America Inc.

5 dispositifs dans la base de données

  • Modèle / numéro de série
    TRACE ITEM SERIAL NUMBERS in the UNITED STATES: US713B0171, US713B0207, US813B0017, US813B0018, US813B0019, US813B0020, US813B0022, US813B0023, US813B0024, US813B0025, US813B0122, US813B0123, US813B0125, US813B0127, US813B0128, US813B0129, US813B0131, US813B0132, US813B0134, US813B0135, US813B0233, US813B0234, US813B0235, US813B0236, US813B0237, US813B0238, US813B0239, US813B0240, US813B0423, US813B0424, US813B0425, US813B0426, US813B0427, US813B0428, US813B0429, US813B0430, US813B0431, US813B0432, US813B0433, US813B0434, US913B0033, US913B0034, US913B0035, US913B0036, US913B0037, US913B0038, US913B0039, US913B0040, US913B0057, US913B0060, US913B0061, US913B0062, US913B0063, US913B0064, US913B0065, US913B0066, US913B0106, US913B0107, US913B0108, US913B0109, US913B0110, US913B0111, US913B0112, US913B0113, US913B0144, US913B0176, US913B0201, US913B0256, US913B0257, US913B0350, US913B0363, US913B0364, US913B0365, US913B0366, US913B0367, US913B0368, US913B0369, US913B0380, US913B0384, US913B0395, US913B0396, US913B0397, US913B0398, US913B0399, US913B0400, US913B0401, US913B0402, US913B0409, US913B0441, US913B0444, US913B0445, US913B0446, US913B0447, US913B0503, US913B0522, US913B0614, US913B0615, US913B0616, US913B0618, US913B0619, and USO13B0043.  TRACE ITEM SERIAL NUMBERS INTERNATIONAL:  US513B0224, US513B0295, US513B0387, US513B0560, US613B0299, US713B0085, US713B0086, US713B0118, US713B0119, US713B0120, US713B0122, US713B0123, US713B0126, US713B0149, US713B0171, US713B0172, US713B0192, US713B0193, US713B0194, US713B0195, US713B0204, US713B0205, US713B0206, US713B0208, US713B0324, US713B0325, US713B0358, US713B0359, US713B0437, US713B0439, US713B0440, US713B0532, US713B0537, US713B0573, US713B0574, US713B0575, US713B0576, US713B0577, US713B0578, US713B0579, US713B0580, US713B0581, US713B0582, US713B0583, US713B0584, US713B0586, US713B0587, US713B0588, US713B0589, US713B0590, US713B0592, US713B0593, US713B0594, US713B0595, US813B0015, US813B0016, US813B0032, US813B0037, US813B0038, US813B0039, US813B0121, US813B0151, US813B0152, US813B0153, US813B0154, US813B0155, US813B0156, US813B0157, US813B0158, US813B0159, US813B0160, US813B0161, US813B0162, US813B0163, US813B0164, US813B0183, US813B0184, US813B0185, US813B0186, US813B0187, US813B0188, US813B0189, US813B0190, US813B0191, US813B0192, US813B0193, US813B0194, US813B0195, US813B0196, US813B0232, US813B0396, US813B0397, US813B0398, US813B0399, US813B0400, US813B0401, US813B0402, US813B0403, US813B0404, US813B0405, US813B0406, US813B0407, US813B0408, US813B0409, US813B0410, US813B0411, US813B0468, US813B0469, US813B0470, US813B0471, US813B0472, US813B0473, US813B0474, US813B0475, US813B0476, US813B0477, US813B0478, US813B0479, US813B0480, US913B0026, US913B0027, US913B0028, US913B0029, US913B0030, US913B0031, US913B0032, US913B0041, US913B0051, US913B0059, US913B0085, US913B0086, US913B0101, US913B0102, US913B0103, US913B0104, US913B0105, US913B0130, US913B0132, US913B0133, US913B0134, US913B0135, US913B0137, US913B0138, US913B0139, US913B0140, US913B0141, US913B0142, US913B0143, US913B0177, US913B0178, US913B0179, US913B0185, US913B0202, US913B0216, US913B0217, US913B0218, US913B0219, US913B0220, US913B0221, US913B0222, US913B0223, US913B0232, US913B0252, US913B0253, US913B0254, US913B0255, US913B0259, US913B0260, US913B0299, US913B0300, US913B0301, US913B0302, US913B0303, US913B0316, US913B0317, US913B0318, US913B0319, US913B0332, US913B0333, US913B0336, US913B0337, US913B0338, US913B0339, US913B0340, US913B0349, US913B0359, US913B0360, US913B0385, US913B0408, US913B0458, US913B0459, US913B0486, US913B0487, US913B0488, US913B0489, US913B0490, US913B0499, US913B0504, US913B0505, US913B0506, US913B0565, US913B0566, US913B0567, US913B0568, USO13B0040, and USO13B0041.
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide Distribution-USA (nationwide) including the states of CT, HI, IL, IN, KY, MA, ME, MI, NC, OR, PA, TX, and WA, and the countries of Australia, Austria, Belgium, Chile, Denmark, Finland, France, Germany, India, Ireland, Republic of Korea, Latvia, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, and United Kingdom.
  • Description du dispositif
    EPIQ 7 Ultrasound System, Model 795200, 795201, 795202. || Intended for diagnostic ultrasound imaging and fluid flow analysis.
  • Modèle / numéro de série
    All systems below software v1.1.2 Which includes software v1 and v 1.1 (All lot codes shipped prior to 2/28/2014).
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide Distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NM, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, and WV; and countries of: Australia, Austria, Belgium, Canada, Chili, Ecuador, Egypt, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Korea Rep, Malaysia, Netherlands, New Zealand, Norway, Peru, Philippines, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom.
  • Description du dispositif
    Philips Model EPIQ 7 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis. || Catalogue numbers 795200 / 795201 / 795202 / 795204 / 795205 / 795206 || EPIQ is intended for diagnostic ultrasound imaging and fluid flow analysis. The collection of Doppler velocity results values are used to assist in diagnosing the existence, location and severity of cardiac or vasculare stenosis.
  • Modèle / numéro de série
    All systems below software v1.1.2 Which includes software v1 and v 1.1 (All lot codes shipped prior to 2/28/2014).
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide Distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NM, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, and WV; and countries of: Australia, Austria, Belgium, Canada, Chili, Ecuador, Egypt, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Korea Rep, Malaysia, Netherlands, New Zealand, Norway, Peru, Philippines, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom.
  • Description du dispositif
    Philips Model EPIQ 5 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis. || Catalogue numbers 795200 / 795201 / 795202 / 795204 / 795205 / 795206 || EPIQ is intended for diagnostic ultrasound imaging and fluid flow analysis. The collection of Doppler velocity results values are used to assist in diagnosing the existence, location and severity of cardiac or vasculare stenosis.
  • Modèle / numéro de série
    Serial numbers of units distributed in the USA: C13A-00592, C13C-00131, C13C-00150, C13C-00270, C13C-00679, C13C-00681, C13C-00769, C13D-00048, C13D-00197, C13E-00088, C13E-00168, C13E-00234, C13E-00242, C13E-00245, C13E-00326, C13E-00340, C13E-00368, C13E-00372, C13E-00380, C13E-00382, C13E-00390, C13E-00406, C13E-00495, C13E-00507, C13E-00511, C13E-00516, C13E-00526, C13E-00537, C13F-00011, C13F-00012, C13F-00013, C13F-00022, C13F-00024, C13F-00027, C13F-00031, C13F-00032, C13F-00033, C13F-00035, C13F-00038, C13F-00043, C13F-00047, C13F-00053, C13F-00054, C13F-00059, C13F-00071, C13F-00077, C13F-00078, C13F-00102, C13F-00104, C13F-00110, C13F-00114, C13F-00118, C13F-00126, C13F-00131, C13F-00137, C13F-00138, C13F-00140, C13F-00143, C13F-00146, C13F-00148, C13F-00168, C13F-00172, C13F-00173, C13F-00174, C13F-00180, C13F-00181, C13F-00198, C13F-00202, C13F-00212, C13F-00214, C13F-00215, C13F-00217, C13F-00218, C13F-00224, C13F-00225, C13F-00227, C13F-00232, C13F-00233, C13F-00236, C13F-00238, C13F-00239, C13F-00240, C13F-00244, C13F-00246, C13F-00250, C13F-00255, C13F-00266, C13F-00273, C13F-00274, C13F-00275, C13F-00278, C13F-00284, C13F-00291, C13F-00294, C13F-00307, C13F-00309, C13F-00318, C13F-00321, C13F-00322, C13F-00323, C13F-00329, C13F-00330, C13F-00331, C13F-00332, C13F-00336, C13F-00348, C13F-00353, C13F-00468, C13F-00543, C13F-00559, C13F-00567, C13F-00578, C13F-00590, C13F-00593, C13F-00595, C13F-00599, C13F-00604, C13F-00607, C13F-00651, C13F-00778, C13G-00008, C13G-00010, C13G-00035, C13G-00056, C13G-00057, C13G-00282, C13G-00284, and C13H-00062.  Serial numbers of units outside the USA:  C13C-00004, C13C-00256, C13C-00305, C13C-00323, C13C-00608, C13C-00639, C13E-00009, C13E-00021, C13E-00036, C13E-00048, C13E-00057, C13E-00058, C13E-00065, C13E-00163, C13E-00250, C13E-00265, C13E-00273, C13E-00311, C13E-00323, C13E-00328, C13E-00339, C13E-00341, C13E-00345, C13E-00346, C13E-00349, C13E-00350, C13E-00351, C13E-00358, C13E-00393, C13E-00395, C13E-00404, C13E-00407, C13E-00409, C13E-00424, C13E-00437, C13E-00441, C13E-00452, C13E-00454, C13E-00455, C13E-00499, C13E-00503, C13E-00518, C13E-00527, C13E-00535, C13E-00548, C13F-00010, C13F-00280, C13F-00286, C13F-00293, C13F-00298, C13F-00302, C13F-00303, C13F-00305, C13F-00306, C13F-00316, C13F-00317, C13F-00343, C13F-00345, C13F-00346, C13F-00358, C13F-00363, C13F-00369, C13F-00371, C13F-00381, C13F-00385, C13F-00391, C13F-00400, C13F-00420, C13F-00424, C13F-00428, C13F-00436, C13F-00473, C13F-00498, C13F-00539, C13F-00565, C13F-00579, C13F-00580, C13F-00583, C13F-00586, C13F-00587, C13F-00606, C13F-00619, C13F-00627, C13F-00629, C13F-00649, C13F-00666, C13F-00668, C13F-00677, C13F-00678, C13F-00682, C13F-00685, C13F-00687, C13F-00726, C13F-00742, C13F-00746, C13F-00748, C13F-00755, C13F-00758, C13F-00760, C13F-00782, C13F-00797, C13F-00815, C13F-00870, C13F-00889, C13F-00896, C13F-00938, C13G-00026, C13G-00039, C13G-00084, C13G-00126, C13G-00132, C13G-00142, C13G-00237, C13G-00247, C13G-00293, C13G-00315, C13G-00331, C13G-00337, C13G-00345, C13G-00407, C13G-00435, C13G-00451, and C13H-00035.
  • Classification du dispositif
  • Classe de dispositif
    3
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide Distribution - US Distribution including the states of AL, CA, CO, WA DC, FL, GA, HI, KS, MD, ME, MI, MS, NC, NM, NY, PA, TX, VA, WA and WI., and the countries of Australia, Austria, Canada, France, Germany, India, Italy, Ireland, Japan, Malaysia, Netherlands, Norway, Switzerland, Taiwan and United Kingdom.
  • Description du dispositif
    Philips HeartStart FR3 Automated External Defibrillator (AED), models 861388 and 861389. || The HeartStart FR3 is a battery powered automated external defibrillator (AED) designed for use by responders who have been trained in Basic Life Support (BLS), Advanced Life Support (ALS), or another physician-authorized emergency medical response program. The HeartStart FR3 is used to treat suspected victims of ventricular fibrillation (VF), the most common cause of sudden cardiac arrest (SCA), and certain ventricular tachycardia (VTs). Both models are used with disposable defibrillator pads applied to potential victims of SCA with the following symptoms: -Unresponsiveness -Absence of normal breathing If in doubt, apply the pads. The HeartStart FR3 is intended for adults and children over 55 pounds (25 kg) or 8 years old. Both models 861388 and 861389 are also intended for children under 55 pounds (25 kg) or 8 years old when used with the optional Infant/Child Key. If the Infant/Child Key is not available, or you are uncertain of the child's age or weight, do not delay treatment.
  • Modèle / numéro de série
  • Classification du dispositif
  • Classe de dispositif
    3
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide Distribution - USA Nationwide
  • Description du dispositif
    Philips HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs). || Product Usage: The models M5066A, M5067A, and M5068A are designed for use on a person in sudden cardiac arrest who is: -Unresponsive when shaken -Not breathing normally If in doubt, apply the pads. A training class in CPR/AED use is recommended for anyone who may use the HS1. If the person is an infant or child younger than eight years old or weighs less than 55 lbs (25 kg), use the special infant/child pads. If the child appears older/larger, use the adult pads. Do not delay treatment to determine the child s exact age or weight. If in doubt, use the adult pads. || Models M5066A (for HS1 Onsite), M5067A (OnSite) , and M5068A (for HS1 Home).