sigmaspectrum VOLUMETRIC INFUSION PUMP with Master Drug Library

  • Modèle / numéro de série
    Contact CDRH for list of affected serial numbers (5976 units)
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    Nationwide Distribution, Puerto Rico and Canada
  • Description du dispositif
    sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique Device Identifier GTIN 00085412091570, Rx ONLY -- Sigma, LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended.
  • Manufacturer
  • 1 Event

Fabricant

  • Adresse du fabricant
    Baxter Healthcare Corp., 1 Baxter Pkwy, Deerfield IL 60015-4625
  • Société-mère du fabricant (2017)
  • Source
    USFDA