Device Recall GE SIGNA Pioneer

  • Modèle / numéro de série
    Mfg Lot or Serial # System ID  001612TA1 N/A 001718TA6 336397P3T  002331TA7 083132425984 001719TA4 250598MR1 001424TA1 M2850013 001537TA0 TP0008MR07 001662TA6 157022MR01 001184TA1 ZC0002 001652TA7 N/A 002227TA7 ZC0003 002278TA0 ZC0004 002330TA9 ZC0006 002529TA6 N/A 001485TA2 34026MRS05 002279TA8 34225MRS02 002395TA2 34185MRS02
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide Distribution - US Nationwide in the states of SC, NC and the countries of Australia, Canada, France. Germany, Japan and Turkey.
  • Description du dispositif
    SIGNA Pioneer MR Systems || Product Usage: || The SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Pioneer reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
  • Manufacturer
  • 1 Event

Fabricant

  • Adresse du fabricant
    GE Medical Systems, LLC, 3000 N Grandview Blvd, Waukesha WI 53188-1615
  • Société-mère du fabricant (2017)
  • Source
    USFDA