Rappel de Cardiolipin IgM ELISA Test System, Product Number 438450CE. The test kit consists of a test plate, and various reagents in individual vials (positive and negative controls, calibrator, diluent, substrate, stop solution, wash buffer.). The individual vials are secured in the card board box. The test kit which is coated with the appropirate antigen, are placed in a foil pouch, which is also placed in the middle of the cardboard box.

Selon U.S. Food and Drug Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à United States qui a été fabriqué par Zeus Scientific Inc.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    29963
  • Classe de risque de l'événement
    Class 3
  • Numéro de l'événement
    Z-0214-05
  • Date de mise en oeuvre de l'événement
    2004-08-30
  • Date de publication de l'événement
    2004-11-20
  • Statut de l'événement
    Terminated
  • Pays de l'événement
  • Date de fin de l'événement
    2006-10-03
  • Source de l'événement
    USFDA
  • URL de la source de l'événement
  • Notes / Alertes
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notes supplémentaires dans les données
    System, Test, Anticardiolipin Immunological - Product Code MID
  • Cause
    During packaging, lyme plates were packaged in this kit in error.
  • Action
    Wampole Laboratories was contacted by Zeus on 8/30/3004. Wampole customers were then contacted by phone and letter to destroy all products on hand.

Device

  • Modèle / numéro de série
    Lot 04027128 exp 2005-10
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    25 units were sent directly from Zeus to their direct account Wampole Laboratories. The distributor contacted their customers to discard all affected kits. 8 units were shipped overseas to Israel. There were no ships to government facilities.
  • Description du dispositif
    Cardiolipin IgM ELISA Test System, Product Number 438450CE. The test kit consists of a test plate, and various reagents in individual vials (positive and negative controls, calibrator, diluent, substrate, stop solution, wash buffer.). The individual vials are secured in the card board box. The test kit which is coated with the appropirate antigen, are placed in a foil pouch, which is also placed in the middle of the cardboard box.
  • Manufacturer

Manufacturer

  • Adresse du fabricant
    Zeus Scientific Inc, 200 Evans Way, Somerville NJ 08876-3767
  • Source
    USFDA