Rappel de Device Recall Baxter FloGard 6300 Dual Channel Volumetric Infusion Pump

Selon U.S. Food and Drug Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à United States qui a été fabriqué par Baxter Healthcare Corp..

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    36872
  • Classe de risque de l'événement
    Class 2
  • Numéro de l'événement
    Z-0635-2007
  • Date de mise en oeuvre de l'événement
    2006-11-27
  • Date de publication de l'événement
    2007-03-20
  • Statut de l'événement
    Terminated
  • Pays de l'événement
  • Date de fin de l'événement
    2010-01-19
  • Source de l'événement
    USFDA
  • URL de la source de l'événement
  • Notes / Alertes
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notes supplémentaires dans les données
    Infusion Pump - Product Code FRN
  • Cause
    There is the potential for non-detection of, and no alarm for, upstream occlusion for flo-gard volumetric infusion pumps, if an occlusion occurs above a flexible chamber such as a drip chamber, buretrol or blood filter.
  • Action
    Baxter sent the 11/27/06 Urgent Device Correction letter to all Flo-Gard Volumetric Infusion Pump customers, to the attention of the Vice President of Nursing and the Biomedical Engineer, via first class mail to alert them to the potential for non-detection of, and no alarm for, upstream occlusion under specific conditions for Flo-Gard Volumetric Infusion Pumps if the occlusion occurs above a flexible chamber (such as a drip chamber, Buretrol or blood filter), and to provide the addendum to the operator manual to address these specific conditions. The letters included the mitigations and risks associated with each of the conditions that could potentially cause this situation. This information was included in the letter, as well as the Operator''s Manual Addendum, which Baxter requested to attach a copy to all Flo-Gard Operator''s Manuals in their facility. Baxter instructed the customers to share this information with all staff members, services and facilities that use Flo-Gard pumps. Any questions were directed to Baxter''s Global Technical Services at 1-800-843-7867, prompt 2. Although Baxter previously notified customers on 3/5/99 that they would no longer support or service the Flo-Gard 6200 and 6300 devices as of 12/1/99, these product codes were included in this letter due to a similar potential for this issue. Should the customer still have these 6200 or 6300 devices in their possession, Baxter recommended that they be taken out of service due to obsolescence of the appropriate components. The accounts were requested complete the enclosed reply form acknowledging receipt of the letter and dissemination of the information to their staff.

Device

  • Modèle / numéro de série
    all serial numbers
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide including USA, Puerto Rico, Argentina, Australia, Bahrain, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Costa Rica, Denmark, Ecuador, Egypt, El Salvador, Guatemala, Hong Kong, Honduras, India, Indonesia, Ireland, Italy, Korea, Lebanon, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Peru, Philippines, Poland, Russia, Singapore, Slovakia, Slovenia, South Africa, Sweden, Taiwan, Thailand, Turkey, United Kingdom and Venezuela.
  • Description du dispositif
    Baxter Flo-Gard 6300 Dual Channel Volumetric Infusion Pump, catalog #2M8048, 2M8048R (refurbished); Baxter Healthcare Corporation, Deerfield, IL 60015
  • Manufacturer

Manufacturer

  • Adresse du fabricant
    Baxter Healthcare Corp., Rt. 120 & Wilson Rd, Round Lake IL 60073
  • Source
    USFDA