Rappel de Device Recall Colleague 3, 3 CX and 3 CXE Volumetric Infusion Pumps

Selon U.S. Food and Drug Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à United States qui a été fabriqué par Baxter Healthcare Corp..

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    38219
  • Classe de risque de l'événement
    Class 1
  • Numéro de l'événement
    Z-1091-2007
  • Date de mise en oeuvre de l'événement
    2007-06-20
  • Date de publication de l'événement
    2007-07-18
  • Statut de l'événement
    Terminated
  • Pays de l'événement
  • Date de fin de l'événement
    2011-12-20
  • Source de l'événement
    USFDA
  • URL de la source de l'événement
  • Notes / Alertes
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notes supplémentaires dans les données
    Infusion Pump - Product Code FRN
  • Cause
    A software anomaly is causing newly upgraded colleague triple channel infusion pumps to alarm, display an error code (16:310:867:0002) and stop the infusion. this occurs during user programming with all three channels simultaneously infusing fluids. in reported cases, the pump stopped infusing and both an audible and a visual alarm notification were activated.
  • Action
    Baxter telephoned the direct accounts on 6/20-21/07, informing them that Baxter has received multiple reports of newly upgraded Colleague Triple Channel Infusion Pumps alarming, displaying an error code and stopping infusion. These reported events occurred during user programming with all three channels simultaneously infusing fluids. In all cases, the pump stopped the infusions and both an audible and visual alarm notification were activated. The accounts were requested to immediately remove the affected triple channel pumps, identified by two attached hangtags entitled "Battery" and "Power On", from service. The accounts were informed that their Baxter representative would be in contact with them to coordinate the replacement of the affected pumps with upgraded single channel pumps or non-upgraded triple channel pumps, if available. Any questions were directed to Baxter's Medication Delivery Services at 1-800-843-7867. Baxter sent follow-up Urgent Product Recall letters dated 6/22/07.

Device

  • Modèle / numéro de série
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide, including USA, Canada, Australia, Belgium, New Zealand, South Africa, Thailand, and Singapore.
  • Description du dispositif
    Baxter Colleague Triple Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in Singapore; product codes 2M8153, 2M8163, 2M9163
  • Manufacturer

Manufacturer

  • Adresse du fabricant
    Baxter Healthcare Corp., Rt. 120 & Wilson Rd, Round Lake IL 60073
  • Source
    USFDA