Rappel de Device Recall Posey

Selon U.S. Food and Drug Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à United States qui a été fabriqué par J T Posey Company.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    49498
  • Classe de risque de l'événement
    Class 2
  • Numéro de l'événement
    Z-0286-2009
  • Date de mise en oeuvre de l'événement
    2008-07-25
  • Date de publication de l'événement
    2008-12-02
  • Statut de l'événement
    Terminated
  • Pays de l'événement
  • Date de fin de l'événement
    2012-03-30
  • Source de l'événement
    USFDA
  • URL de la source de l'événement
  • Notes / Alertes
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notes supplémentaires dans les données
    restraint - Product Code FMQ
  • Cause
    Failure to restrain: this recall was initiated after a complaint was received by the firm regarding a posey 2217sl unit indicating that a patient was able to pull the rivet or washer through the synthetic leather allowing the patient to have freedom of arm movement. a second complaint was received indicating that a patient was able to tear the strap at the adjustment notch on a posey 221 7sl unit.
  • Action
    On July 29, 2008 Posey sent a Recall Letter to all affected customers. The recall letters were sent by certified mail with return receipt signature. The recall letter instructed the customers not to distribute or use the affected products. Any existing inventory subject to this recall was requested to be returned to Posey Company. Also, if they were a distributor, they were asked to please additionally forward a copy of the recall notice and reply form to each of their customers to whom you have further distributed the subject product. A Recall Fax Back form was provided with the recall letter for the affected facilities to fax back to inform Posey of their existing inventory. An RMA number was issured to each facility and a return UPS label was provided for any product that will be returned to Posey. Additionally, the sales representatives for each of the customers will be contacted by Posey Customer Service.

Device

  • Modèle / numéro de série
    Catalog Numbers: 2217SL and 2217B ; Note: All 22l7SL and 2217B units are affected by this recall.
  • Classification du dispositif
  • Classe de dispositif
    1
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide, USA, Australia, Canada, France & The Netherlands.
  • Description du dispositif
    Posey Synthetic Leather and Biothane Waist and Wrist Restraints, Catalog Numbers: 2217SL and 2217B