Rappel de Locking Tear-Away Sheath Introducer Product Line. || ACCUGUIDE SINGLE LUMEN 4FR PICC P460K 5/; || P560K PICC 5FR W/INTRO KIT; || MTI451845 4.5FR MINI T/A; || MTI551845 5.5FR MINI TA; || MTI651845 6.5FR MINI T/A; || P460OTW SNGL LUM PICC W/MTI; || P4552K DBL LUM PICC INTRO KIT; || P5552K DBL LUM PICC INTRO KIT; || P4552KOTW DBL LUM OTW PICC INTRO KIT LF; || LOCKING T/A INTRODUCERS 11FR; || INTRADYN TEAR-AWAY INTRODUCER 4F; || INTRADYN TEAR-AWAY INTRODUCER 5F; || INTRADYN TEAR-AWAY INTRODUCER 6F; || INTRADYN TEAR-AWAY INTRODUCER 7F; || INTRADYN TEAR-AWAY INTRODUCER 8F; || INTRADYN TEAR-AWAY INTRODUCER 9F; || INTRADYN TEAR-AWAY INTRODUCER 10F; || INTRADYN TEAR-AWAY INTRODUCER 11F; || INTRADYN TEAR-AWAY INTRODUCER 12F; || INTRADYN TEAR-AWAY INTRODUCER 15F; || INTRADYN TEAR-AWAY INTRODUCER 16F; || 23HI120 12F X 23CM HEMO INTRO; || 23HI130 13F X 23CM HEMO INTRO; || 30STA8 8FR X30CM T/A INTRO -NJ*; || 30STA9 9FR X 30CM SHEATH -NJ*; || 30STA10 10FR X 30CM T/A INTRO -NJ*; || 30STA12 12FR X 30CM T/A INTRO -NJ*; || 30STA16 16FR X 30CM SHEATH; || BOSTON SCI 16F X 9.54 STEP DIL/SHEATH; || 45STA9 45CM T/A INTRODUCER -NJ*; || LUTHER 6F - T/A ASSM CLIP; || 10HI14038 14F X 11CM HEMO W/WI; || BOSTON SCI 12FX 21.35 DIL/SHEATH; || 10HI12038 12F X 11CM HEMO W/WI; || 10HI13038 13F X 11CM HEMO W/WI; || BOSTON SCI 9F X 9.54 STEP DIL/SH; || BOSTON SCI 9F X 15.44 STEP DIL/SH; || BOSTON SCI 9F X 21.35 STEP DIL/SH; || BOSTON SCI 11F X 9.54 STEP DIL/SH; || BOSTON SCI 11F X 15.44 STEP DIL/SH; || BOSTON SCI 12F X 9.54 DIL/SHEATH; || BOSTON SCI 12F X 15.44 DIL/SHEATH; || BOSTON SCI 14F X 9.54 DIL/SHEATH; || BOSTON SCI 14F X 15.44 DIL/SHEATH; || BOSTON SCI 16F X15.44 DIL/SHEATH; || 30STA15 15FR LONG TA INTRODUCR; || 30STA11 11FR LONG TA INTRODUCR -NJ*; || CARDIAC PACEMAKER T/A SHEATH; || 14FR TA DILATOR/SHEATH-BULK; || 12FR INTRODUCER TEAR AWAY BNS; || 12FR TA DIL/SHEATH LC; || 8FR TA SHEATH; || 10FR TA SHEATH; || 16FR TA SHEATH; || 8/10FR LOCK T/A INTRO STEPPED; || 6FR T/A DILATOR/SHEATH; || 6FR LOCKING T/A INTRO ASSY; || 14FR X 143MM LOCKING T/A INTRODUCERS; || PTFE 4.5 FR BARD ACCESS INTRODUCERS; || PTFE 6.0 FR BARD ACCESS INTRODUCERS; || 7F LOCKING INTRO SYSTEM; || 8 (O/S) FR. LOCKING TEARAWAY INTRODUCER; || 9 FR (0/S) LOCKING INTRO SYSTEM; || 12F LOCKING INTRO SYSTEM; || 14FR LOCKING T/A INTRODUCER; || 10FR X 18 CM LOCKING T/A INTRODUCER; || 14FR X 12MM LOCKING T/A INTRODUCER; || 10.5FR X 18CM LOCKING T/A INTRODUCER; || 10FR X 23CM LOCKING T/A INTRODUCER; || 10FR LOCKING TEAR-AWAY INTRODUCERS; || INTRADYN TEAR-AWAY INTRODUCER 10.5F; || 30STA14 14FR X 30CM T/A INTRO -NJ*; || BOSTON SCIENTIFIC 15FR LOCKING TEARAWAY; || CELSA 5FR LOCKING T/A INTRODUCER; || CORDIS 4'' BLUNT TIP T/A SHEATH; || MEDTRONIC PS MED INTRO ASM 14F -NJ*; || MEDTRONIC PS MED INTRO ASM 15F.

Selon U.S. Food and Drug Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à United States qui a été fabriqué par B. Braun Medical, Inc..

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    32210
  • Classe de risque de l'événement
    Class 2
  • Numéro de l'événement
    Z-1062-05
  • Date de mise en oeuvre de l'événement
    2005-05-27
  • Date de publication de l'événement
    2005-08-03
  • Statut de l'événement
    Terminated
  • Pays de l'événement
  • Date de fin de l'événement
    2007-01-03
  • Source de l'événement
    USFDA
  • URL de la source de l'événement
  • Notes / Alertes
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notes supplémentaires dans les données
    Introducer, Catheter - Product Code DYB
  • Cause
    Reports that the tearaway introducer sheaths have cracked hubs/handles and/or improper peeling of the sheath during use.
  • Action
    Recall letters shere sent out to the firm''s customers on 5/27/2005 via US Mail, Return Receipt Requested.

Device

  • Modèle / numéro de série
    Catalog #, Lot Numbers:  333603--Lot #s 0060529076,0060542787, 0060551433, 0060552115, 0060553837, 0060562411, 0060702019, 0060702049. 333604--0060553838, 0060708443, 0060713806. 333606--0060472888, 0060516047, 0060583876, 0060710305, 0060710306, 0060751880. 333608--0060569225, 0060703806, 0060729030, 3360734585. 333609--0060703808. 333610--0060738852. 333612--0060542791, 0060549251, 0060553963, 0060702064, 0060708037. 333613--0060549252, 0060553909, 0060702069, 0060702071, 0060702074. 333614--0060521374, 0060551766, 0060702082. 610064--0060709374. 610100--0060702177, 0060761533. 610101--0060702179. 610102--0060710159. 610103--0060713850. 610104--0060734898, 0060741005. 610105--0060722688, 0060740847. 610106--0060702186. 610108--0060702192. 610109--0060713856. 610112--0060719604. 610113--0060723081. 610145--0060570243. 610146--0060560608. 610150--0060551068. 610151--0060707659, 0060710792. 610152--0060710794. 610153--0060551073, 0060568835. 610155--0060708367, 0060708367, 0060742089. 610177--0060558435, 0060707652, 0060709662, 0060746191, 0060766191. 610264--0060707534. 610407--0060732173. 610559--0060523821, 0060555811, 0060572434, 0060709663, 0060726953, 0060733425. 610570--0060726955. 610652--0060707368, 0060737064. 610654--006545559, 0060552179, 0060744541. 610811--0060473107, 0060473108, 0060560316, 0060738144. 610812--0060726992. 610813--0060757144. 610814--0060572400, 0060732297. 610815--0060538361, 0060542288, 0060757084. 610816--0060520549, 0060523748, 0060538364, 0060572403, 0060710820, 0060722731. 610817--0060544240, 0060545503, 0060569903, 0060571978. 610818--0060528048, 0060561924, 0060709739, 0060726994. 610819--0060545504, 0060709740, 0060727002. 610820--0060707651. 610830--0060712212, 0060742386. 610843--0060709735. 610894--0060544824, 0060545508, 0060557624, 0060707071, 0060715230, 0060728009, 0060752044, 0060752045. 610916--0060710874, 0060710875. 610923--0060497968, 0060545511, 0060754189, 0060516182, 0060545512, 0060569910. 610974--0060719256. 610976--0060517487, 0060551012, 0060564301, 0060571977, 0060709065, 0060719972. 610982--0060725523. 611109--0060523767, 0060545524, 0060551019, 0060555703, 0060706362, 0060707365, 0060709725, 0060716210. 611172--0060706038. 611840--0060519177, 0060540542, 0060545484, 0060552193, 0060564379, 0060707285, 0060711192, 0060722451. 612001--0060528366, 0060541715, 0060709669. 612122--0060752209. 612124--0060755565, 0060759158. 614006--0060545527, 0060545529, 0060548273, 0060555801, 0060564933, 0060569911, 0060572409, 0060706138, 0060706158, 0060706169, 0060706174, 0060709679, 0060749179. 614007--0060473083, 0060494998, 0060503536, 0060507452, 0060517491, 0060545530, 0060545531, 0060551054, 0060564304, 0060564934, 0060566303, 0060566304, 0060566305, 0060712277, 0060720315, 0060724332, 0060726782, 0060750181. 614008--0060545533, 0060548274, 0060554892, 0060555802, 0060564936, 0060568832, 0060569912, 0060572414, 0060576479, 0060576483, 0060576484, 0060712278, 0060712279, 0060712280, 0060737734. 614012--0060747064. 614037--0060547187, 0060735870. 614100--0060709646, 0060709647, 0060715276, 0060720258, 0060521185, 0060536288, 0060538205, 0060551057, 0060560474, 0060572417. 614101--0060709664, 0060709667, 0060715113. 614102--0060562423, 0060710927. 614104--0060580188, 0060707545. 614400--0060588319, 0060707536, 0060707537, 0060707538, 0060712281, 0060722450, 00607727033, 0060749069. 610107A--0060702191, 0060746124. 610154A--0060709726. 7B2807--0060709702, 0060709703, 0060709704, 0060709705, 0060709709, 0060709710, 0060709711, 0060709719, 0060749170. 7C2100--0060707500, 0060707501, 0060709659. 7C6311--0060482936, 0060489182, 0060496750, 0060505790, 0060506316, 0060523417, 006055541633, 0060543824, 0060543825, 0060543826, 0060548258, 0060555712, 0060557290, 0060561310, 0060578814, 0060710970, 0060720149, 0060726624. 7M3882--0060551059, 0060572130, 0060710976, 0060738147, 0060746766. 7M3883--0060707663, 0060725281.
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    The firm distributed the product to 235 customers both US and International. The international customers are located in Germany, Canada, India, Japan, France, Venezuela, Ireland, and Chile.
  • Description du dispositif
    Locking Tear-Away Sheath Introducer Product Line. || ACCUGUIDE SINGLE LUMEN 4FR PICC P460K 5/; || P560K PICC 5FR W/INTRO KIT; || MTI451845 4.5FR MINI T/A; || MTI551845 5.5FR MINI TA; || MTI651845 6.5FR MINI T/A; || P460OTW SNGL LUM PICC W/MTI; || P4552K DBL LUM PICC INTRO KIT; || P5552K DBL LUM PICC INTRO KIT; || P4552KOTW DBL LUM OTW PICC INTRO KIT LF; || LOCKING T/A INTRODUCERS 11FR; || INTRADYN TEAR-AWAY INTRODUCER 4F; || INTRADYN TEAR-AWAY INTRODUCER 5F; || INTRADYN TEAR-AWAY INTRODUCER 6F; || INTRADYN TEAR-AWAY INTRODUCER 7F; || INTRADYN TEAR-AWAY INTRODUCER 8F; || INTRADYN TEAR-AWAY INTRODUCER 9F; || INTRADYN TEAR-AWAY INTRODUCER 10F; || INTRADYN TEAR-AWAY INTRODUCER 11F; || INTRADYN TEAR-AWAY INTRODUCER 12F; || INTRADYN TEAR-AWAY INTRODUCER 15F; || INTRADYN TEAR-AWAY INTRODUCER 16F; || 23HI120 12F X 23CM HEMO INTRO; || 23HI130 13F X 23CM HEMO INTRO; || 30STA8 8FR X30CM T/A INTRO -NJ*; || 30STA9 9FR X 30CM SHEATH -NJ*; || 30STA10 10FR X 30CM T/A INTRO -NJ*; || 30STA12 12FR X 30CM T/A INTRO -NJ*; || 30STA16 16FR X 30CM SHEATH; || BOSTON SCI 16F X 9.54 STEP DIL/SHEATH; || 45STA9 45CM T/A INTRODUCER -NJ*; || LUTHER 6F - T/A ASSM CLIP; || 10HI14038 14F X 11CM HEMO W/WI; || BOSTON SCI 12FX 21.35 DIL/SHEATH; || 10HI12038 12F X 11CM HEMO W/WI; || 10HI13038 13F X 11CM HEMO W/WI; || BOSTON SCI 9F X 9.54 STEP DIL/SH; || BOSTON SCI 9F X 15.44 STEP DIL/SH; || BOSTON SCI 9F X 21.35 STEP DIL/SH; || BOSTON SCI 11F X 9.54 STEP DIL/SH; || BOSTON SCI 11F X 15.44 STEP DIL/SH; || BOSTON SCI 12F X 9.54 DIL/SHEATH; || BOSTON SCI 12F X 15.44 DIL/SHEATH; || BOSTON SCI 14F X 9.54 DIL/SHEATH; || BOSTON SCI 14F X 15.44 DIL/SHEATH; || BOSTON SCI 16F X15.44 DIL/SHEATH; || 30STA15 15FR LONG TA INTRODUCR; || 30STA11 11FR LONG TA INTRODUCR -NJ*; || CARDIAC PACEMAKER T/A SHEATH; || 14FR TA DILATOR/SHEATH-BULK; || 12FR INTRODUCER TEAR AWAY BNS; || 12FR TA DIL/SHEATH LC; || 8FR TA SHEATH; || 10FR TA SHEATH; || 16FR TA SHEATH; || 8/10FR LOCK T/A INTRO STEPPED; || 6FR T/A DILATOR/SHEATH; || 6FR LOCKING T/A INTRO ASSY; || 14FR X 143MM LOCKING T/A INTRODUCERS; || PTFE 4.5 FR BARD ACCESS INTRODUCERS; || PTFE 6.0 FR BARD ACCESS INTRODUCERS; || 7F LOCKING INTRO SYSTEM; || 8 (O/S) FR. LOCKING TEARAWAY INTRODUCER; || 9 FR (0/S) LOCKING INTRO SYSTEM; || 12F LOCKING INTRO SYSTEM; || 14FR LOCKING T/A INTRODUCER; || 10FR X 18 CM LOCKING T/A INTRODUCER; || 14FR X 12MM LOCKING T/A INTRODUCER; || 10.5FR X 18CM LOCKING T/A INTRODUCER; || 10FR X 23CM LOCKING T/A INTRODUCER; || 10FR LOCKING TEAR-AWAY INTRODUCERS; || INTRADYN TEAR-AWAY INTRODUCER 10.5F; || 30STA14 14FR X 30CM T/A INTRO -NJ*; || BOSTON SCIENTIFIC 15FR LOCKING TEARAWAY; || CELSA 5FR LOCKING T/A INTRODUCER; || CORDIS 4'' BLUNT TIP T/A SHEATH; || MEDTRONIC PS MED INTRO ASM 14F -NJ*; || MEDTRONIC PS MED INTRO ASM 15F.
  • Manufacturer

Manufacturer

  • Adresse du fabricant
    B. Braun Medical, Inc., 901 Marcon Blvd, Suite 200, Allentown PA 18109-9512