Ortho-Clinical Diagnostics

14 dispositifs dans la base de données

  • Modèle / numéro de série
    Catalog/REF Nos. 166 8409 and 829 7749. All lots.
  • Classification du dispositif
  • Classe de dispositif
    1
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide distribution
  • Description du dispositif
    VITROS¿ Chemistry Products LIPA Slides GEN 48, Coating 3235, REF/Catalog No. 166 8409 (60 slides per cartridge and 5 cartridges per box - 300 slides per box) and REF/Catalog No. 829 7749 (18 slides per cartridge and 5 cartridges per box - 90 slides per box). LIPA Slides are processed by higher volume (or mainframe) VITROS¿ Chemistry Systems typically used in hospital laboratories. Ortho-Clinical Diagnostics, a Johnson & Johnson company. Ortho-Clinical Diagnostics, Inc. Rochester, NY 14626. 510(k) #: K845027 (18 January 1985). Shelf life: Cat. # 166 8409: 12/01/07, Cat. # 829 7749: 12/01/07-1/01/08
  • Modèle / numéro de série
    Kit Lot 2430 (exp. 11/16/11), Kit Lot 2440 (exp. 01/08/12), Kit Lot 2450 (03/09/12), Kit Lot 2471 (exp. 7/14/12)
  • Classification du dispositif
  • Classe de dispositif
    2
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide Distribution - USA, including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
  • Description du dispositif
    VITROS¿ Chemistry Products Calibrator Kit 24, Catalog No. 680 1708. Common/Usual Name: VITROS Calibrator Kit 24. For in vitro diagnostics use only. VITROS Chemistry Products Calibrator Kit 24 is used to calibrate the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of albumin in urine. Ortho-Clinical Diagnostics Inc.
  • Modèle / numéro de série
    enGenTM Laboratory Automation Systems wiht TCAutomation Software Versions 3.1.1 and below.  Parallel Bypass for Vitros 5.1 FS AT (VITROS 5, 1 FS Chemistry System, VITROS 5600 Integrated System and CITROS 3600 Immunodiagnostics System use this bypass)  Parallel Bypass for Vitros 950  Perpendicular Bypass for Vitros 250  Perpendicular Bypass for Vitros 5, 1 FS AT  Parallel Bypass for Advia Centaur  Parallel Bypass for Abbott Architect i2000
  • Classification du dispositif
  • Classe de dispositif
    1
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide distribution: USA and countries including: Canada, Chile, England, France, Germany, Italy, Mexico, Singapore and Spain.
  • Description du dispositif
    The enGenTM Laboratory Automation System consists of 3 main components including Instrument Management Subsystem (Instrument Manager), Automated Track Subsystem (TCAutomation) and Configuration File developed by Ortho-Clinical Diagnostics (OCD). The Instrument Manager is a Windows-based software product that operates on a Pentium-class (or higher) PC processor. The instrument management subsystem is capable of connectivity with clinical analyzers, a Laboratory Information Systems (LIS), barcode label printers and other third party devices. The TCAutomation is a pre-analytical and post-analytical sample management system. The configuration file enables operational support of the site specific layout and instrument selection. || The VITROS 5, 1 FS Chemistry System with enGenTM Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric Immunoassay methods.)
  • Modèle / numéro de série
    Serial Numbers between 30003000 and 30005177; All VITROS¿ ECi/ECiQ systems that have Software Version 3.8 or below installed and on which the VITROS¿ B-hCG II (human chorionic gonadotropin) Reagent Packs is processed are affected by this correction. --- Note: All systems that are installed worldwide have Software Version 3.8 or lower installed on them, but B-hCG is not processed on all systems worldwide.
  • Classification du dispositif
  • Classe de dispositif
    1
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide Distribution - USA (nationwide) and including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
  • Description du dispositif
    VITROS¿ ECiQ Immunodiagnostic System, Catalog No. 1922814 --- COMMON/USUAL NAME: VITROS¿ ECiQ System --- Ortho-Clinical Diagnostics, 100 Indigo Creek Drive, Rochester, NY 14626 --- All systems that have Software Version 3.8 or below installed and on which the VITROS¿ Total B-hCG II Reagent Pack (human chorionic gonadotropin) is processed. || Product Usage: || For in vitro diagnostic use only. To perform random access, batch and STAT immunodiagnostic assays on human fluid specimens
  • Modèle / numéro de série
    Serial Numbers between 30000127 and 30002743. All VITROS¿ ECi/ECiQ systems that have Software Version 3.8 or below installed and on which the VITROS¿ Total B-hCG II (human chorionic gonadotropin) Reagent Packs is processed are affected by this correction. --- Note: All systems that are installed worldwide have Software Version 3.8 or lower installed on them, but Total B-hCG is not processed on all systems worldwide.
  • Classification du dispositif
  • Classe de dispositif
    1
  • Dispositif implanté ?
    No
  • Distribution
    Worldwide Distribution - USA (nationwide) and including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
  • Description du dispositif
    VITROS¿ ECi Immunodiagnostic System, Catalog No. 8633893 --- COMMON/USUAL NAME: VITROS¿ ECi System -- Ortho-Clinical Diagnostics, 100 Indigo Creek Drive, Rochester, NY 14626 --- All systems that have Software Version 3.8 or below installed and on which the VITROS¿ Total B-hCG II Reagent Pack (human chorionic gonadotropin) is processed. || Product Usage: || For in vitro diagnostic use only. To perform random access, batch and STAT immunodiagnostic assays on human fluid specimens
9 en plus

53 fabricants avec un nom similaire

En savoir plus sur les données ici

48 en plus