(a) Catalog Number PCVJG5, Lot Numbers: 73E1700794, 73G1700271, 73G1700515, 73H1700081, 73H1700367 (b) Catalog Number PCVMD5, Lot Numbers: 73E1700796, 73G1700273, 73H1700083 (c) Catalog Number PCVHCA5, Lot Numbers: 73E1700798, 73H1700084 (d) Catalog Number PCVGG5, Lot Numbers: 73E1700793, 73G1700272, 73H1700082, 73H1700247 (e) Catalog Number PCVSC5, Lot Number: 73H1700569
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Australia, Belgium, Hong Kong, Japan, S. Korea and Singapore.
Description du dispositif
Teleflex(R) Percuvance(TM) Percutaneous Surgical System, Components: || (a) 5 mm Atraumatic Johans Grasper, Catalog Number PCVJG5 || (b) 5 mm Maryland Dissector, Catalog Number PCVMD5 || (c) 5 mm Hem-o-lok Clip Applier, Catalog Number PCVHCA5 || (d) 5 mm Traumatic Gripper Grasper, Catalog Number PCVGG5 || (e) 5 mm Scissors, Catalog Number PCVSC5 || The Percuvance Percutaneous Surgical System is a family of instruments used to perform laparoscopic procedures. Without the need for additional insertion conduits (trocars), the 2.9mm insulated Shafts are inserted percutaneously with the assistance of an Introducer needle tool tip. The Shaft is then advanced retrograde through a 5mm, or larger, port to facilitate the extracorporeal attachment of interchangeable Tool Tips. These interchangeable tools can be used to grasp, retract, manipulate, cut, coagulate and deliver Hem-o-lok¿ ligating clips to soft tissue. Each Percuvance System Handle includes an integrated male cautery adapter that can be connected to most monopolar electrosurgical units (ESU) via a standard 4mm monopolar cable.
Worldwide Distribution - US including AL, AZ, ARK, CA, CO, CT, FL, GA. HI, KS, KY, IL, IN, IA, LA, MD, MA, MI, MN, MS, MO, MT, NE, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV and Internationally to Canada, Chile, & Japan
Description du dispositif
Sheridan/CF, Cuffed Tracheal Tube, Catalog # 5-10114 and Sheridan/HVT Cuffed Tracheal Tube, Catalog # 5-10315 || A tracheal tube is inserted into a patients mouth or nose for airway management
Worldwide distribution. U.S. (Nationwide) and Internationally to Canada, Belgium, Singapore and Thailand.
Description du dispositif
LMA MADdy, Rx only, || Product Usage: || These products are used for the delivery of topical anesthesia via an atomized spray to the oral, nasal, pharyngeal or laryngeal mucosa. Teleflex Medical is recalling these products as they may produce a straight stream instead of a fully atomized plume of medication. It is unlikely that serious adverse health consequences will occur in the event of a failure to deliver an atomized plume; however, this may result in inadequate topical anesthesia which may lead to some discomfort, further attempts to deliver topical anesthesia, or the use of alternative methods of anesthesia.